Bio Safety Cabinets
Biological Safety Cabinets are partly recirculating laminar airflow enclosures with high efficiency particulate air (HEPA) filtration of exhaust air and an air barrier at the work opening. HEPA-filtered vertical laminar airflow which is recirculated in the work zone creates an ultra-clean work environment for product protection. An air barrier between the operator and the work zone is maintained by a flow of room air into a full width grille in the work opening. The barrier air mixes with the laminar flow air in a sump underneath the work surface, and is exhausted to the outside environment via a HEPA filter and an optional virus burn-out. All positive pressure zones and filter seals are surrounded by negative pressure zones, so as to contain potentially hazardous aerosols.
Material of construction made of waterproof commercial ply with exterior covered by industrial laminated sheets & interiors coated with synthetic rubber paint. Side & back panels covered by lamination. Working Table Made of perforated stainless steel - satin finished Front door Made of clear Perspex sheet - vertical sliding
| HEPA FILTER |
Media Glass fiber paper - imported |
| Particle retention |
0.3 micron |
| Efficiency |
99.97 % |
| Pressure drop |
23 mm WG |
| Size - at supply |
48" x 24" x 6" 1 Number |
| Size - at Exhaust |
12" x12" x 6" 1 Number |
| PRE FILTER |
Media Non woven - synthetic polyester |
| Particle retention |
10 - 15 micron |
| Efficiency |
90 % |
| Pressure drop |
6 mm WG |
| Size |
800 x 400 x 60 mm |
| Blower Assembly |
By 1/3 HP , single phase, 1440 RPM motor , directly coupled with a pair of aluminum, centrifugal impeller . Capacity 1000 CFM, Pressure 30 mm WG |
| Exhaust Assembly |
By ¼ HP , Single phase, 1440 RPM motor , directly coupled with a aluminum impeller . Capacity - 300 CFM, Ducting by PVC pipe with rain guard - DIA - 150 mm |
| Ultra violet lamp |
3 feet, 30 Watt - PHILIPS - HOLLAND make - 1 No |
| Illumination |
4 feet, 40 Watt - with diffusers |
| Standard accessories |
Pressure manometer , Additional power point, Gas inlet nozzle, floor leveling screws, wire chord |
| Power supply |
220 V, single phase, 50 Hz |
|